Monocyto-ID Listeria Scan

Call for Price: +30 26810 88000

Ref.No.: FX100025 – 25 vials, FX100100 – 100 vials

Guaranteed safe checkout

p_s_01 p_s_04p_s_03 p_s_02
Certified Goods Personal Consultant A Large Assortment

Need Help? Call Us +30 26810 88000
Monday - Friday 9:00 - 17:00

Description

Standard method for biochemical detection and identification of Listeria monocytogenes

Description
Listeriosis is a serious infection, classified as a zoonosis, which results from the ingestion of food contaminated with the bacterium Listeria monocytogenes. It leads to a severe clinical picture in infants, immunocompromised individuals and poses significant risks to foetuses, while pregnant women usually have mild symptoms.

Contents
FX100025 – 25 vials R
FX100100 – 100 vials R
Note: All kit components are stable until the expiration date on the label. Protect them from light and contamination during use. Do not use the reagents after the expiration date. Store reagents at 2-8°C.
Shelf life: 12 months from date of production
Storage and stability: 2-8°C

Samples
Food, milk, dairy products, fish
Note: for sample preparation, please contact the scientific support department of MenidiMedica
Biotech

Procedure
– Incubation of the sample in suitable nutrient medium (e.g. Ottaviani Agosti Agar Base) for 24 hours at 37⁰C in an incubator
– Place a vial – found on the package – with the Listeria monocytogenes biochemical detection and identification reagent in the incubator at the same time as the incubation described in step a is performed.
– Remove the medium and the vial from the incubator.
– Open the vial and add one drop or 50 ul of deionised water to the vial
– Collect a formed colony from the nutrient medium with a swab and dip it into the vial, shaking gently until the contents of the swab are dissolved in the vial.
– Incubate the vial in the incubator at 37⁰C for 6 hours
– Remove the vial from the incubator and check the colour that has appeared

Interpretation
Green: positive sample for Listeria monocytogenes
Sensitivity: 100% for Listeria monocytogenes
Specificity: 100% for Listeria monocytogenes
A comparison study with other methods was performed on a total number of samples n=500.

Monocyto-ID Listeria Scan PDF ENG